If any undissolved solids are present in your sample, they must be removed before a satisfactory spectrum can be obtained. 

-the spinner tachometer
will register 0 or some very large(<100) number), and
Place your sample in the turbine with the same careful 'inching' procedure and adjust the tube's position in the turbine with the spinner gauge attached to the magnet. The bottom of the tube should extend all the way to the bottom of the spinner gauge, but no further. Be very careful to position the tube properly - if it is too far into the turbine, it may break when placed in the magnet.
- parallel to the floor). The sample should begin to spin as indicated by a reading on the spinner tachometer. Turn the spin control clockwise until the arrow on the knob is at the 2 o'clock position. The tachometer reading should stabilize somewhere between 15 and 60 RPS. If it does not spin or the spin rate is very erratic, consult your instructor or Dr. Badger.